- Use a low-cost microcontroller and can be operated without a personal computer.
- Fast data acquisition and analysis, showing results in 30 seconds.
- Low cost, low power, and low maintenance.

Block Diagram of QMIA
07T-1110209
Applicable for hand-held immunoassay readers (AS docket number: 07A-1100702) to enable rapid and quantitative testing of both antigens and antibodies.
Huan-Cheng Chang; Hui Yuen Yung; Yi-Xiu Tang; Shui-Yuan Wu
If you are interested in this technology, please fill in the information on the webpage under the manufacturer option (Request for Technology Licensing, Material Transfer, or Sponsored Research) on the website of this office

. Data processing algorithm for QMIA. The upper panel is the raw experimental data, while the lower panel is the signal after second-order differentiation. The labels “C” and “T” denote the control line and test line, respectively. Note that the baseline of the original signal is not flat, which prevents precise area measurement for the absorption bands of both control and test lines. A smoothing filter (such as the Savitzky-Golay smoothing filter) is thus applied to the experimental data, followed by the second-order differentiation of the data as shown in the lower panel. This allows easy determination for the end points of both absorption bands and hence their area integration for the quantification purpose。

Hardware, Firmware, and Software Designs for Quantitative Mobile ImmunoAnalyzer (QMIA)